Location: Home > IEC > IEC 62467-1:2009 pdfダウンロード

IEC 62467-1:2009 pdfダウンロード

IEC 08-28
IEC 62467-1:2009 pdfダウンロード

IEC 62467-1:2009 pdfダウンロード。MEDICAL ELECTRICAL EQUIPMENT – DOSIMETRIC INSTRUMENTS AS USED IN BRACHYTHERAPY – Part 1: Instruments based on well-type ionization chambers
1 Scope and object
This part of IEC 62467 specifies the performance and some related constructional requirements of WELL – TYPE IONIZATION CHAMBERS and associated measurement apparatus, as defined in Clause 3, intended for the determination of a quantity, such as AIR KERMA STRENGTH or REFERENCE AIR KERMA RATE in photon radiation fields or ABSORBED DOSE TO WATER at a depth, in photon and beta radiation fields used in BRACHYTHERAPY , after appropriate calibration for a given type of source.
This International Standard covers the techniques for the quantification of the quantity appropriate for the BRACHYTHERAPY source under consideration. This quantity may be AIR KERMA STRENGTH or REFERENCE AIR KERMA RATE at 1 m, or ABSORBED DOSE TO WATER at a depth (e.g. 2 mm or 5 mm). Measurement of these quantities may be accomplished by a variety of WELL – TYPE IONIZATION CHAMBERS or systems currently available for this purpose. This standard applies to products intended for low dose rate, high dose rate, intravascular, both photon and beta, BRACHYTHERAPY measurements.
It does not apply to instruments for nuclear medicine applications. The application of the standard is limited to instruments that incorporate WELL – TYPE IONIZATION CHAMBERS as detectors. The intended use is the measurement of the output of radioactive, encapsulated sources for intracavitary (insertion into body cavities) or interstitial (insertion into body tissue) applications. The object of this standard is a) to establish requirements for a satisfactory level of performance for WELL – TYPE CHAMBER SYSTEMS , and b) to standardize the methods for the determination of compliance with this level of performance.
This standard is not concerned with the safety aspects of WELL – TYPE CHAMBER SYSTEMS . The WELL – TYPE CHAMBER SYSTEMS covered by this standard are not intended for use in patient environment. The electrical safety of WELL – TYPE CHAMBER SYSTEMS is covered in IEC 61 01 0-1 . The operation of the electrometer measuring system is covered in IEC 60731 .
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
IEC 60050-393:2003, International Electrotechnical Vocabulary – Part 393: Nuclear instrumentation – Physical phenomena and basic concepts
IEC 6041 7, Graphical symbols for use on equipment
IEC 60580:2003, Medical electrical equipment – Dose area product meters

Download